You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 19 Next »

Attention:

Please note that the system can only be operated upon request. Operations of the Salinometer requires experienced scientific staff on board. The system is not operated by the ship's crew. Announce operations with the Salinometer to AWI-Logistics prior to the cruise and clarify the data transfer after the cruise.

Summary

The Precision Salinometer calculates the salinity of a water sample from the measurement of the electrical conductivity and the determination of its temperature. The calculations are based on the Practical Salinity Scale 1978 (PSS78) as published by UNESCO.

Manufacturer Optimare
Model Precision Salinometer
Serial No. n/a
Type salinity sensor
SENSOR-Link

Contacts

Name Institution Role
Sandra Tippenhauer Alfred Wegener Institute, Helmholtz Centre for Polar and Marine Research Engineer In Charge, Data Provider

Components

The Salinometer contains two controlled water baths (pre-bath 0.5l and main bath 13l) each equipped with Peltier module, stirrer, heat adapter and temperature sensor (SBE3 plus). Peltier element and stirrer are mechanically matched and located at one side of the bath. The main bath additionally contains the conductivity cell (modified SBE4). The system is connected via intake, tubing, filling pump and probe outlet. The Salinometer is equipped with hard disk for data storage, all software included, USB and Ethernet Network connections and a touch screen as user interface, optional an USB keyboard or mouse can be connected. The conductivity cell can and should be calibrated ('standardized', i.e. performing a one point calibration) prior to a measurement using IAPSO Standard Seawater. The SBE3 plus can be returned to manufacturer Sea-Bird Electronics for recalibration.

Data logging, storage and archiving

Logged parameters

Conductivity, Salinity, Bath Temperature (incl. drift)

Central geographical ship's position and time standard

No time synchronising needed.
No online ship's position needed.

Rawdata storage on board

Local data storage. Operator is in charge of data backup.

File format(s)

ASCII

Calibration certificates

YES

Location of certificates

Sandra Tippenhauer

Documentation

  • No labels