...
Note |
---|
|
Two models of Nevzorov Probes exist (CWCMB2 and CWCMU4) which operate after the same principle. Each system consists of a control box and a sensor head. Operation of the Nevzorov probes requires experienced staff on board the aircraft. |
Summary
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Summary |
---|
|
The Sonde consits of two seperate sensors Sky Tech Nevzorov-probe is an aircraft probe designed for measurements of LWC ( liquid water content) and TWC ( total (ice plus liquid plus ice) water content (LWC and TWC) in of clouds and fog fogs in the range of 0.003 g/m3 to 3 g/m3. The Nevzorov-probe belongs to the group of so-called "hot wire constant temperature probes". A sensor unit consists of two sensors: collector and reference. The principle of operation is based on separate measurements of power needed to compensate heat losses caused by evaporation of cloud water and convective losses on the collector sensor and power needed to compensate convective heat losses on a reference sensor. Liquid and total water content are measured by separate LWC and TWC sensors, mounted on the same sensor vane. |
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Metadata |
---|
|
Manufacturer | SkyTech Research Limited |
---|
Model | CWCMB2 and CWCMU4 |
---|
Serial No. | n/a |
---|
Type |
---|
not assigned |
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Images |
---|
|
|
...
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Contacts |
---|
|
| Name | Institution | RoleNo contacts available.
---|
Available devices and Specifications
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Subdevices |
---|
|
Type | Serial Number | REGISTRY-Link |
---|
Nevzorov Probe | water content analyzerwater content analyzer |
Specifications of
...
Nevzorov Probe (without mounting)
Metadata list |
---|
|| Dimension Control Box (L x W x H)| 350 cm x 390 x 50 mm ||
|| Weight Control Box | 4.3 kg||
|| Power Supply| 28 VDC ||
|| Data interface | analog (V)|| |
|
Position
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | LocalFrame |
---|
|
Origin Description:Info | The sensor head of the Nevzorov Probe is installed in the nose compartment of the aircraft. The electronic control unit is installed in a rack in the aircraft cabin. | X | 0.0 none (no description given) | Y | 0.0 none (no description given) | Z | 0.0 none (no description given) |
|
Data logging, storage and archiving
...
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Parameters |
---|
|
Sensor O2ARegistry Output Type | Unit |
---|
liquid water content | water content | g/cm3 | ice + liquid (total) water content | water content | g/cm3 |
|
Rawdata storage on board
Metadata (Metadata Plugin) |
---|
|
During flight LWC and TWC data are stored within the aircraft onboard DMS system. |
|
Data archiving on land
Metadata (Metadata Plugin) |
---|
|
After the flight LWC and TWC data are accessible via the land system of the aircraft data managment system (DMS). |
|
DMS
Metadata list |
|| Module name| B809 Nevzorov Sonde, B438 Nevzorov Sonde||
|| Device name | Nevzorov Probe||
|| Parameters | ErrorText; LWC Probe \[V\]; Status; TWC Probe \[V\] ||
|
---|
|
Documentation
Sensorweb integration |
---|
urn | aircraft:bt-67:nev |
---|
contenttype | Resources |
---|
|
|
Also see https://sensor.awi.de/ for calibration certificates of the sensors.