Thermosalinograph data stored in data acquisition system DShip are uncorrected measurement data.
Summary
The thermosalinograph is an instrument used to accurately determine sea surface conductivity and temperature. The remote temperature sensor SBE38 is integrated with the SBE21 Thermosalinograph.
Manufacturer | Seabird |
---|---|
Model | SBE21 with SBE38 |
Serial No. | n/a |
Type | thermosalinograph |
REGISTRY-Link | ↪ REGISTRY (8561) |
Contacts
Name | Institution | Role |
---|---|---|
Sandra Tippenhauer | Alfred Wegener Institute for Polar and Marine Research | Editor |
FIELAX GmbH | FIELAX Gesellschaft für wissenschaftliche Datenverarbeitung mbH | Engineer In Charge |
Components
The thermosalinograph consists of the SBE21, the external thermometer SBE38, the NMEA interface box and a pipe system with a flow meter. The system is located in the bow room of the tank deck. Two devices of type SBE21 and two devices of type SBE38 exist on RV Heincke. They are used in rotation and calibrated regularly.
Name | Model | Serial Number | REGISTRY-Link |
---|---|---|---|
Temperature Sensor SBE38 | SBE38 | 3853181-0474 | ↪ REGISTRY (8645) |
Thermosalinograph SBE21 | SBE21 | 2153181-3334 | ↪ REGISTRY (8558) |
Position
Info | no xyz-position given.The device is installed in the bow room of the tank deck. |
X | 0.0 (no unit given) |
Y | 0.0 (no unit given) |
Z | 0.0 (no unit given) |
Data logging, storage and archiving
Logged parameter
Parameter | O2ARegistry Output Type | Unit |
---|---|---|
conductivity | conductivity | Hz |
internal temperature | internal temperature | Hz |
external temperature | water temperature | Hz |
first voltage | voltage | V |
secondary voltage | voltage | V |
Central geographical ship's position and time standard
Rawdata storage on board
DShip
Device Name |
Thermosalinograph |
---|---|
Parameters |
Conductivity (S/m) |
Temperature internal (°C)|
Data archiving on land