Please note that the system can only be operated upon request. Operations with the MSCL-S requires experienced scientific staff on board. The system is not operated by the ship's crew. Announce operations with the MSCL-S to AWI-Logistics prior to the cruise and clarify the data transfer after the cruise.

Summary

Multi-Sensor Core Loggers capture physical measurands of sediment cores automated and non-destructive.


ManufacturerGeotek Ltd.
ModelMSCL-S
Serial NoNEU
TypeMulti-Sensor Core Logger

Contacts

RoleName
Principal InvestigatorFrank Niessen
Data ScientistCatalina Gebhardt
Engineer in ChargeDietmar Penshorn

Components

The system consits of the measuring bench with a ballscrew, which pushes core section past 6 to 7 sensors measuring Gamma Density, P-wave Velocity, Magnetic Susceptibility, Point Magnetic Susceptibility, Colour Spectrophotometry, Linescan Imaging and optional Natural Gamma Spectrometry. All sensors simultaneously acquire the data, which is displayed real-time during logging. The acquired data is stored locally in the electronics with simple user interface.
|| Sensor Type |  Gamma Density Sensor | P-Wave System | Magnetic Susceptibility Loop Sensor MS2C | Magnetic Susceptibility Point Sensor MS2E | CM-700D Spectrophotometer | Geoscan V Linescan Imaging | optional: Natural Gamma Spectrometry ||
|| Manufacturer | Geotek Ltd. | Geotek Ltd. | Bartington Instruments| Bartington Instruments | Konica Minolta Measuring Instruments (Konica) | Geotek Ltd. | Geotek Ltd. ||
|| Components | Gamma ray source (137-Caesium source with energies at 0.662 MeV) \\ detector \\ sensor stand mount (alignment of source/detector with center of core) | Geotek ultrasonic (230 kHz) acoustic transducers \\ Laser micrometer \\ PRT temperature Probe | Magnetic Susceptibility Loop Sensor MS2C | Magnetic Susceptibility Point Sensor MS2E | CM-700D Spectrophotometer | Geoscan V Linescan Imaging Camera specifications: \\  Pixel Sensors: 1x 5K CGD \\ Visible or Ultraviolet light source \\ ADC resolution: 14 bits per colour channel \\ correlated double sampling with exposure control \\ max pixel rate: 3.2M pixels/sec \\ max scan rate: 200 lines/sec \\ down core resolution: upt ot 10 microns \\ camera control: auto focus, aperture and lighting \\ data transmission: gigabyte ethernet interface \\ Canon lens mount | three 3"x3" NaI(Tl) detectors  with photomultiplier tube and Multi-Channel Analyser (MCA) \\ detector housings: 6" diameter lead shields \\ interface and electronics supplied in integrated 19" rack unit ||

Data logging, storage and archiving

Logged parameters

Gamma Density, P-wave Velocity, Magnetic Susceptibility (Loop), Magnetic Susceptibility (Point), Colour Spectrophotometry, Geoscan V Linescan Imaging, optional: Natural Gamma Spectrometry

Rawdata storage on board

Local storage on MSCL-S electronic. Operator is in charge of data backup.

Documentation