Note |
---|
|
Please note that the system can only be operated upon request. Operations with the MSCL-S NG requires experienced scientific staff on board. The system is not operated by the ship's crew. Announce operations with the MSCL-S NG to AWI-Logistics prior to the cruise and clarify the data transfer after the cruise. |
Summary
Metadata (Metadata Plugin) |
---|
|
Multi-Sensor Core Logger - Natural Gamma System is a stand-alone natural gamma system designed to obtain both total and full spectral natural gamma data. For high resolution and quality spectral data from sedimentary rocks the system is configured with three NaI(Tl) detectors. |
...
Metadata (Metadata Plugin) |
---|
|
*Manufacturer*: Geotek Ltd.
*Model*: MSCL-NG
*Serial No*:
*Type*: Multi-Sensor Core Logger-Natural Gamma System |
Responsible Persons
Role | Name |
---|
Engineer in Charge | Dietmar Penshorn |
Data Scientist | Frank Niessen |
Components
Metadata (Metadata Plugin) |
---|
|
The sensor comprises three 3"x3" NaI(Tl) detectors housed in 6" diameter lead shields. Each detector uses a NaI(Tl) crystal optically coupled to a photomultiplier tube and a Multi-Channel Analyser (MCA). The detectors make measurements over 1024 channels calibrated to correspond to 0-3 MeV energy range. The complete system is controlled through custom-designed Geotek MSCL software, which enables either multi-channel spectral measurements or simply total counts to be recorded. The software gives the user full control of sampling positions and intervals down the core and control of count times. All the electronics are supplied in a single integrated 19" rack unit that contains of all the necessary external electronics including the motor control unit, the microprocessor and rack mounted PC. |
Metadata (Metadata Plugin) |
---|
|
|| Sensor Type | MSCL-NG Natural Gamma System||
|| Manufacturer | Geotek Ltd. ||
|| Components | three 3"x3" NaI(Tl) detectors with photomultiplier tube and Multi-Channel Analyser (MCA) \\ detector housings: 6" diameter lead shields \\ interface and electronics supplied in integrated 19" rack unit ||
|
Position
No fix onboard location.
Data logging, storage and archiving
Logged parameters
total and full spectral natural gamma data
Central geographical ship's position and time standard
Metadata (Metadata Plugin) |
---|
|
No time synchronising. |
Metadata (Metadata Plugin) |
---|
|
No online ship's position. |
Rawdata storage on board
Metadata (Metadata Plugin) |
---|
|
Local storage on MSCL-NG electronic. Operator is in charge of data backup.
|
Dship
Metadata (Metadata Plugin) |
---|
|
No data transfer to DShip. |
Data archiving on land
Metadata (Metadata Plugin) |
---|
|
Not implemented, responibility with respective user. |
Documentation
http://www.geotek.co.uk/wp-content/uploads/2017/06/Natural_Gamma.pdf
...