Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.
Comment: Automatically updated by Registry2Confluence-Transporter (polarstern@ingestsrv1, ask o2a-support@awi.de for questions)

...

Note
titleAttention:

Please note that the system can only be operated upon request. Operations of the Salinometer requires experienced scientific staff on board. The system is not operated by the ship's crew. Announce operations with the Salinometer to AWI-Logistics prior to the cruise and clarify the data transfer after the cruise.

Summary

Sensorweb integration
urn vessel:polarstern:ops_ps
contenttype
Metadata (Metadata Plugin)
0Summary

The

Precision

Salinometer

calculates

the

salinity

of

a

water

sample

from

the

measurement

of

the

electrical

conductivity

and

the

determination

of

its

temperature.

The

calculations

are

based

on

the

Practical

Salinity

Scale

1978

(PSS78)

as

published

by

UNESCO.

Image Removed


...

Sensorweb integration
urn vessel:polarstern:ops_ps
contenttypeMetadata
Manufacturer
Optimare
Model

...

Precision Salinometer
Serial No

...

Type: Precision Salinometer

Responsible Persons

...

Components

Metadata (Metadata Plugin)
0Components
The Salinometer contains two controlled water baths (pre-bath 0.5l and main bath 13l) each equipped with Peltier module, stirrer, heat adapter and temperature sensor (SBE3 plus). Peltier element and stirrer are mechanically matched and located at one side of the bath. The main bath additionally contains the conductivity cell (modified SBE4). The system is connected via intake, tubing, filling pump and probe outlet. The Salinometer is equipped with hard disk for data storage, all software included, USB and Ethernet Network connections and a touch screen as user interface, optional an USB keyboard or mouse can be connected. The conductivity cell can and should be calibrated ('standardized', i.e. performing a one point calibration) prior to a measurement using IAPSO Standard Seawater. The SBE3 plus can be returned to manufacturer Sea-Bird Electronics for recalibration.

Position

...

.n/a
Typesalinity sensor
REGISTRY-Link↪ REGISTRY (8445)


Sensorweb integration
urn vessel:polarstern:ops_ps
contenttypeImages

Image Added

Contacts

Sensorweb integration
urn vessel:polarstern:ops_ps
contenttypeContacts
Name Institution Role
Sandra TippenhauerAlfred Wegener Institute, Helmholtz Centre for Polar and Marine ResearchEngineer In Charge, Data Provider

Components

Metadata (Metadata Plugin)
0Components

Data logging, storage and archiving

Logged parameters

Conductivity, Salinity, Bath Temperature (incl. drift)

...

No online ship's position needed.
Metadata (Metadata Plugin)
0timestandard
No time synchronising needed.
Metadata (Metadata Plugin)
0shipsposition

Rawdata storage on board

Metadata (Metadata Plugin)
0Rawdata
Local data storage. Operator is in charge of data backup.

Dship

Metadata (Metadata Plugin)
0DShip
No data transfer to DShip.

Data archiving on land

Metadata (Metadata Plugin)
0DataOnLand
Not implemented, responsibility with respective user.
Metadata list
|| File format(s) | ??? || 
|| Calibration certificates | YES ||
|| Location of certificates | Gereon Budeus ||

Documentation


Metadata list

Documentation

Sensorweb integration
urn vessel:polarstern:ops_ps
contenttypeResources

...