titleAttention:

In case of a system failure just reboot the computer (switch on). The data acquisition program will start automatically.

Summary

...

Sensorweb integration

Components

Metadata (Metadata Plugin)
0Components
All components are in the Aerologie lab (A-108). The system consits of 2 sensor boxes, a data acquisiton system, GPS antenna and a notebook (Linux, Xubuntu 16.04).
In each sensor box is a scintillator, which emits scintillation light when charged particles transmit. The light is transmitted to the photosensor (silicium-photomultiplier by Hamamatsu) via optical fiber. A HV module within the box converts the 5V input power supply to the supply voltage of 70V for the photomultiplier. The signal of the photomultiplier is transmitted to the data acquisition unit. The data acquisition unit needs a 15V power supply, provides the 5V power supply for the sensor boxes, amplifies and links the signals from both boxes, produces and transfers data files via USB cable to the notebook, and faciltiates an input port for signal transmission from the GPS antenna.
The data acquisition unit and the GPS box are connected via cable. The antenna is mounted at the railing above the starboard outside door.
urnvessel:polarstern:myon_detector
contenttypeSummary

The device detects cosmic particles (mainly myons) which are produced by the reaction of cosmic radiation with atomic nuclei or molecules in the upper atmosphere. The system consists of 2 sensor boxes, a data aquisiton system, GPS antenna and a netbook.

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeMetadata
ManufacturerDESY
Model
Serial No.n/a
Typesmall scale facility
REGISTRY-Link↪ REGISTRY (1421)
Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeImages

Image Added Image Added

Contacts

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeContacts
Name Institution Role
Michael WalterDeutsche Elektronen-Synchrotron DESYPrincipal Investigator

Components

Metadata (Metadata Plugin)
0Components

Position

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeLocalFrame
Infono xyz-position given, device located on deck A in room A112
X0.0 (no unit given)
(no description given)
Y0.0 (no unit given)
(no description given)
Z0.0 (no unit given)
(no description given)

...

Data logging, storage and archiving

Logged parameters

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeParameters
Parameter O2ARegistry Output Type Unit
amount of cosmic particlesamountnumber

Central geographical ship's position and time standard

Metadata (Metadata Plugin)
0timestandard
No online time needed for data acquisition. Time stamp needed to identify ship's position and weather data for offline processing.
No online position needed for data acquisition. Offline data processing is conducted with ship's position and weather data via time stamp.

Rawdata storage on board

Raw data is stored in ASCII files on SD memory card and USB stick connected to the notebook. A LAN cable connects the notbook to the data network of Polarstern.
Metadata (Metadata Plugin)
0Rawdata

Data

...

Dship

Metadata (Metadata Plugin)
0DShip
No data transfer to DShip.

Data archiving on land

Metadata (Metadata Plugin)
0DataOnLand
Data is archived at the DESY: http://cosmicatweb.desy.de/ctplot/ 

Documentation

The manuals are located in the folder at the detector on board.

...

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeResources