Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.
Comment: Automatically updated by Registry2Confluence-Transporter (aircrafts@ingestsrv1, ask o2a-support@awi.de for questions)

...

Note
titleAttention:

Two models of Nevzorov Probes exist (CWCMB2 and CWCMU4) which operate after the same principle. Each system consists of a control box and a sensor head.

Operation of the Nevzorov probes requires experienced staff on board the aircraft.

Summary

Sensorweb integration
urnaircraft:bt-67:nev
contenttypeSummary

The Sky Tech Nevzorov-probe is an aircraft probe designed for measurements of liquid and total (ice plus liquid) water content (LWC and TWC) of clouds and fogs in the range of 0.003 g/m3 to 3 g/m3. The Nevzorov-probe belongs to the group of so-called "hot wire constant temperature probes". A sensor unit consists of two sensors: collector and reference. The principle of operation is based on separate measurements of power needed to compensate heat losses caused by evaporation of cloud water and convective losses on the collector sensor and power needed to compensate convective heat losses on a reference sensor. Liquid and total water content are measured by separate LWC and TWC sensors, mounted on the same sensor vane.

Sensor-Info: aircraft:bt-67:nev, 6951


Sensorweb integration
urnaircraft:bt-67:nev
contenttypeMetadata
ManufacturerSkyTech Research Limited
ModelCWCMB2 and CWCMU4
Serial No.n/a
Typewater content analyzer
REGISTRY-Link↪ REGISTRY (6951)


Sensorweb integration
urnaircraft:bt-67:nev
contenttypeImages

...

Sensorweb integration
urnaircraft:bt-67:nev
contenttypeContacts

No contacts available.


Available devices and Specifications

Sensorweb integration
urnaircraft:bt-67:nev
contenttypeSubdevices
Name Model Serial Number REGISTRY-Link
Nevzorov ProbeCWCM-B2n/a↪ REGISTRY (6952)
Nevzorov ProbeCWCM-U4.1n/a↪ REGISTRY (6954)


Specifications of Nevzorov Probe (without mounting)

|| Dimension Control Box (L x W x H)| 350 cm x 390 x 50 mm || || Weight Control Box | 4.3 kg|| || Power Supply| 28 VDC || || Data interface | analogue data (V)||
Metadata list

Position

Sensorweb integration
urnaircraft:bt-67:nev
contenttypeLocalFrame
Origin Description:InfoThe sensor head of the Nevzorov Probe is installed in the nose compartment of the aircraft. The electronic control unit is installed in a rack in the aircraft cabin.
X0.0 none
(no description given)
Y0.0 none
(no description given)
Z0.0 none
(no description given)


Data logging, storage and archiving

...

Sensorweb integration
urnaircraft:bt-67:nev
contenttypeParameters
Parameter Sensor O2ARegistry Output Type Unit
liquid water contentwater contentg/cm3
ice + liquid (total) water contentwater contentg/cm3

...

Metadata (Metadata Plugin)
0Rawdata
During flight LWC and TWC data are stored within the aircraft onboard DMS system.

Data archiving on land

After the flight LWC and TWC data are accessible via the land system of the aircraft data managment system (
Metadata (Metadata Plugin)
0DataOnLand

DMS

...

DMS

|| Module name| B809 Nevzorov Sonde, B438 Nevzorov Sonde|| || Device name | Nevzorov Probe|| || Parameters | ErrorText; LWC Probe \[V\]; Status; TWC Probe \[V\] ||
Metadata list


Documentation

Sensorweb integration
urnaircraft:bt-67:nev
contenttypeResources

...