Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeSummary

The device detects cosmic particles (mainly myons) which are produced by the reaction of cosmic radiation with atomic nuclei or molecules in the upper atmosphere. The system consists of 2 sensor boxes, a data aquisiton system, GPS antenna and a netbook.

vessel:polarstern:myon_detector (1421)

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeMetadata
ManufacturerDESY
Model
Serial No.n/a
Typesmall scale facility

...

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeContacts
Name Institution Role
Michael WalterDESYPrincipal Investigator
polarstern dshipAlfred-Wegener-Institutedship connector

Components

Metadata (Metadata Plugin)
0Components
All components are in the Aerologie lab (A-108). The system consits of 2 sensor boxes, a data acquisiton system, GPS antenna and a notebook (Linux, Xubuntu 16.04).
In each sensor box is a scintillator, which emits scintillation light when charged particles transmit. The light is transmitted to the photosensor (silicium-photomultiplier by Hamamatsu) via optical fiber. A HV module within the box converts the 5V input power supply to the supply voltage of 70V for the photomultiplier. The signal of the photomultiplier is transmitted to the data acquisition unit. The data acquisition unit needs a 15V power supply, provides the 5V power supply for the sensor boxes, amplifies and links the signals from both boxes, produces and transfers data files via USB cable to the notebook, and faciltiates an input port for signal transmission from the GPS antenna.
The data acquisition unit and the GPS box are connected via cable. The antenna is mounted at the railing above the starboard outside door.

...

Sensorweb integration
urnvessel:polarstern:myon_detector
contenttypeLocalFrame
Origin Description:Infono xyz-position given, device located on deck A in room A112

...